-40%

Ground Resistance Tester Soil Resistivity Meter 2000Ω Earth Voltage Measurment

$ 142.56

Availability: 100 in stock
  • Condition: New
  • Ground Resistance Resolution: 0.01Ω
  • Short Circuit Test Current: AC 10mA max
  • Item must be returned within: 30 Days
  • All returns accepted: Returns Accepted
  • Current Range: /
  • Ground Resistance Range: 0.010Ω~2000Ω
  • Accuracy:: ±2%rdg±3dgt
  • Return shipping will be paid by: Buyer
  • MPN: resistance tester
  • Warranty: 1 year
  • Restocking Fee: No
  • Refund will be given as: Money Back
  • Country/Region of Manufacture: China
  • Resolution:: 0.01Ω
  • Data storage:: 400 sets
  • Simple Test Wire: Red 1.6m
  • Brand: VTSYIQI

    Description

    Ground Resistance Tester Soil Resistivity Meter 2000Ω Earth Voltage Measurment
    Instruction

    Digital Earth Resistance Tester specially designed and manufactured for the field measurement of ground resistance, adopts digital and micro-processing technology, 3 wire or 2 wire method to measure ground resistance, with unique calibration function for wire resistance, anti-interference ability and environmental adaptability, to ensure the long-term measurement of high precision, high stability and reliability; The instrument out shell adopts waterproof protection case, anti-collision, anti-fall, waterproof (protection grade IP65),which is strong, safe and durable, the instrument is also equipped with large capacity rechargeable lithium battery pack, especially suitable for outdoor construction site to use.It is widely used in measuring ground resistance of electric power, telecommunication, meteorology, oil field, construction, lightning protection and industrial electrical equipment.
    Digital Earth Resistance Tester has the unique function of wire resistance calibration, and it is more accurate to measure the low value ground resistance in the field, which can avoid the error caused by the change of wire resistance in the test line for a long time usage.It can avoid errors caused by incomplete insertion of test line into instrument interface or bad contact.It can avoid the error caused by the user to change or lengthen the test line..
    Digital Grounding Resistance Tester is composed of host machine, monitoring software, test wires, communication wires etc. The large LCD display of host machine, with white backlight and bar graph indicating that can be seen clearly. At the same time it can store 400 sets of data, realize historical inquiry and online real-time monitoring by monitoring software, dynamic display, with the indicator of max/min/average value, and with alarm value setting and  alarm indicator functions, and with the functions as historical data access, reading, preservation, report forms, printing and so on.
    Product Main Function
    1. Earth resistance measurement, Range: 0.010Ω~2000Ω
    2. Resolution: 0.01Ω
    3. Accuracy: ±2%rdg±3dgt
    4. Earth voltage measurement,Range: AC 0

    600V
    5. Resolution:1V
    6. Data storage: 400 sets
    7. Alarm function: audible and visual alarm
    Specification

    Measurement Range and Accuracy
    Measurement Function
    Range
    Accuracy
    Resolution
    Earth Resistance
    0.01Ω

    20Ω
    ±2%rdg±3dgt
    0.01Ω
    0.1Ω

    200Ω
    (Auxiliary grounding resistance 100Ω±5%, Votage to ground

    10V)
    0.1Ω


    2000Ω

    Earth Voltage
    0

    600V AC
    ±2%rdg±3dgt
    1V

    Note

    23

    ±5
    ℃,
    below75%rh

    Technical Specifications
    Function
    Measurement of earth resistance, voltage to earth
    Power Supply
    DC 9V Alkaline dry cell LR14 1.5VX6pcs, continues working 300 hours

    battery not included

    Backlight
    Controllable screen backlight, suitable for dim places
    Measuring Mode
    Precision 3-wire measurement, simple 2-wire measurement
    Measuring Method
    Earth Resistance: rated current change-pole method, measurement current 20mA Max
    Earth Voltage: Average rectification
    Test Frequency
    500Hz
    Short Circuit Test Current
    AC 10mA max
    Data Mode
    Average value, maximum value, minimum value
    Wire Resistance Verification
    To avoid errors caused by incomplete insertion of test wire into instrument interface or poor contact or replacement of extended test wire by users, so as to make the grounding resistance measurement more accurate
    Display Mode
    4-digital super-large LCD display, white screen backlight
    Measure Indicator
    LED flash indicator and LCD count down display in measurement
    LCD Dimension
    128mm×75mm; display area:124mmX67mm
    Meter Dimension
    215mm(L)×178mm(W)×83mm(H)
    Test Wires
    3 wires: each for red 15m, yellow 10m, and green 5m
    Simple Test Wire
    Red 1.6m
    Auxiliary Grounding Rod
    2PCS: Φ10mm×150mm
    Measuring Rate
    Earth Voltage: about 3 times/second; Earth resistance: about 5 seconds/time
    Measuring Times
    Over 5000 times (20Ω gear measure 10Ω, interval time should be at least 30 seconds)
    Line Voltage
    Measurement of earth voltage: below AC 600V
    Data Storage
    400 sets, “FULL” symbol flash display storage is full
    Data Hold
    Data hold function: “HOLD” symbol display
    Data Access
    Data read function: “READ” symbol display
    Overflow Display
    Exceed measuring range overflow function: “OL” symbol display
    Alarm Function
    Measuring value exceeds alarm setting value, LCD flashes display alarm symbol
    Battery Voltage
    While battery voltage decreases to around 7.2V±0.1V, will display battery voltage low symbol“   ”, and remind to charging the battery
    Power Consumption
    Backlight: 25mA Max
    Standby: 25mA Max(Backlight shut down )
    Measurement: 200mA Max(Backlight shut off)
    Weight
    Tester: 1.325kg
    Test wires: 0.847kg(include the simple test wires)
    Auxiliary Grounding Rod:0.468kg(2pcs)
    Tester package: 0.915kg
    Working Temperature & Humidity
    -10

    -40

    , below 80%rh
    Storage temperature & humidity
    -20

    -60

    , below 70%rh
    Protection Level
    IP65(close the case)
    Overload Protection
    Measure earth resistance: between each interfaces of C(H)-E

    P(S)-ES, AC 280V/3 seconds
    Insulation Resistance
    Over 10MΩ (between circuit and outside shell is 500V)
    Withstanding Voltage
    AC 3700V/rms. (Between circuit and outside shell)
    Electromagnetic Features
    IEC61010-4-3,wireless frequency electromagnetic field≤1V/m
    Protection Type
    IEC61010-1 (CAT

    300V

    CAT IV 150V

    Pollution 2);
    IEC61010-031;
    IEC61557-1 (Earth resistance);
    Packing
    Package
    Outer package
    Carton
    Inner package
    Fabric Bag
    Accessory
    Meter
    1 PCS
    Cloth  Bags
    1 PCS
    Auxiliary grounding rod
    2 PCS
    Test Wires
    4 PCS
    Simple Test Wires
    2 PCS
    Manual/Warranty Card/Certificate
    1 PCS